Electrolyte Analyzer MTN 2OOOC
Solenoid valve-controlled gas-liquid flushing (pipeline anti-clogging). RS232 interface for central computer connection. Operates at 5-40℃, ≤85%RH, 220V±22V/50Hz.
100μL sample volume (whole blood/plasma/serum), ≤1min measurement time. 24h continuous operation, fault self-diagnosis, one/two-point auto-calibration, built-in thermal printer.
Direct measurement via ion-selective electrodes (ISE). Dual-parameter calibration, automatic electrode drift tracking, ensuring accurate results for K⁺/Na⁺/Cl⁻/Ca⁺⁺ etc.
Technical Specifications
- Direct measurement using ion selective electrodes.
The solenoid valve controls the thorough flushing of gas-liquid sections,and the pipeline is simple and not easily blocked.
Automatic tracking of electrode status eliminates the problem of electrode drift.
Dual parameter calibration ensures more accurate sample measurement.
Specially designed quality control testing and statistical functions, easy to use.
24-hour continuous startup with fault self diagnosis function, allowing for
sample testing at any time. One point/two point automatic calibration and
the ability to process quality control data.
Technical parameter
Test project:K+、Na+、Cl-、Ca++、PH、nCa++、TCa++
measuring method:Ion selective electrode (ISE) direct method.
Sample type:Whole blood, plasma, and serum.
Sample size:100μl
Washing method:By using an electromagnetic valve for rapid flushing of
gas-liquid segments, the electrode pipeline is thoroughly
cleaned and not easily clogged.
Measurement time:Not exceeding 1 minute (taking the total balance time measured, including suction, rinsing, and printing).
Display:Large screen, full Chinese display, human-machine dialogue, backlit Chinese LCD display.
Printer:Built in high-speed low-noise thermal printer.
Interface:Standard RS232 interface, capable of connecting to a central computer.
Usage conditions
Power Supply:-220V±22V,50±1Hz
Ambient temperature:5℃-40℃;
Relative humidity:≤85%RH