Benchtop Centrifuge JTGF 80-1
-
This centrifuge utilizes centrifugal separation technology to separate spherical heterogeneous mixtures, suitable for biochemical research such as cell collection and DNA extraction.
-
It has a centrifugal capacity of 20ml×6, a rotational speed of 400 – 4000 rpm, a relative centrifugal force of 1790×g, and a motor power of 25W.
-
The centrifuge operates on a 220V 50Hz power supply and supports a timing range of 1 – 60 minutes or continuous operation (ON). Ensure it is placed on a stable surface, load tubes symmetrically with equal volumes, and always start/stop with the speed control knob at minimum.
Description
I. Overview
Centrifugal separation technology is a new technology developed based on the state of particles in practical centrifugal conditions. Particles with different densities, sizes, or shapes settle at different rates under centrifugal forces of varying speeds, allowing roughly spherical heterogeneous mixtures to be separated by centrifugation. This centrifuge is designed for further biochemical research and large-scale substance separation, such as cell collection, plasma separation, extraction of DNA and proteins from purified preparations, as well as isolation of viruses, large-scale Escherichia coli, cellular components, and nucleoprotein particles.
II. Main Technical Parameters
-
Power Supply: AC 220V 50Hz
-
Motor Power: 25W
-
Rotating Speed: 400–4000 rpm
-
Centrifugal Capacity: 20ml×6
-
Timing Range: 1–60 minutes or continuous operation (ON)
-
Relative Centrifugal Force (RCF): 1790×g
III. Operating Instructions
-
The instrument should be placed on a stable workbench in a clean, tidy environment with moderate temperature, humidity, and good ventilation.
-
Before use, set the speed control knob to the minimum position.
-
Insert centrifuge tubes containing substances to be separated into the instrument’s dedicated holders. To ensure stable operation and avoid excessive vibration, the volume of substances in each tube must be approximately equal, and tubes should be inserted symmetrically. Close the cover.
-
Select the timing mode by turning the timer knob to “Timing” or “Continuous (ON)” position.
-
Connect the power supply, switch the power to “ON”—the indicator light will illuminate. Slowly adjust the speed control knob to the desired rotating speed.
-
Before shutting down, set the speed control knob to the minimum position, turn off the power switch, and disconnect the power supply.