Specific Protein Analyzer HS QR-2000
Dual-platform (nephelometry + turbidimetry), 3-optics & multi-wavelength. Avoids HOOK effect, auto HCT correction, 200T/H speed.
Progressive track sampling, 360° barcode scan, mechanical blood mixing. Supports LIS, emergency insertion, batch detection.
Whole blood/serum/plasma/urine compatible. Multi – stage cleaning (no cross – contamination), reagent auto – dilution, traceable to standards.
Technical Specifications
The HPA-2000 Automatic Specific Protein Analyzer adopts a dual-platform of nephelometry and turbidimetry, based on the three-optics principle and multi-wavelength detection. It effectively avoids the HOOK effect and achieves accurate quantitative analysis. It supports continuous loading, mechanical arm-assisted shaking of whole blood samples, 360° barcode scanning of sample tubes, track-based online operation, and integration into production lines for more efficient detection. It covers multiple detection indicators including inflammation, glucose metabolism, rheumatism, and kidney disease, providing efficient support for medical institutions at all levels.
Efficient Automation and Online Sampling
Progressive track-based sampling for batch detection, supporting online sampling;
Handheld + automatic 360° barcode scanning of sample tubes;
Automatic shaking of whole blood samples supported.
Safety Protection and Multi-stage Cleaning
No sample pretreatment required, with original tube puncture sampling;
Sampling needles equipped with liquid detection and anti-collision protection functions;
Independent multi-stage cleaning to avoid cross-contamination and efficiently save cleaning solution.
Dual-platform of Nephelometry + Turbidimetry
Dual-platform of nephelometry/turbidimetry, three-optics principle, and multi-wavelength detection;
Wide detection range, high sensitivity, and accurate and reliable results;
Detectable sample types: whole blood, serum, plasma, urine;
Rich detection menu with free combination of same-type items, suitable for early screening and treatment detection of various diseases.
Hematocrit (HCT) Correction Function
Automatic correction function to enhance the anti-interference ability of the detection system;
Use hematocrit to correct whole blood detection results, maintaining high consistency with serum detection results.
Simple and Efficient Reagent Detection
Automatic monitoring and dilution to solve the problem of antigen excess;
Wide detection linear range to reduce the probability of sample retesting;
Traceable to national standard products, with high result accuracy.
PARAMETER ITEM | DETAILS | PARAMETER ITEM | DETAILS |
|---|---|---|---|
Detection Principle | Nephelometric immunoassay, Turbidimetry | Ambient Temperature Reagent Positions | 16 positions |
Detection Speed | 200T/H | Blood Mixing | Mechanical arm simulates manual mixing to effectively prevent blood sedimentation |
Sample Positions | 5 racks * 10 positions | Barcode Scanning | Automatic scanning, handheld scanning |
Reaction Positions | 50 cup positions (recyclable) | Emergency Function | Capable of inserting emergency samples at any time, with independent emergency positions |
Sample Types | Whole blood, serum, plasma, urine | Communication Function | Single-direction LIS, Two-way LIS |
Cleaning System | Four-stage cleaning (cleaning solution + purified water combination) | Dimensions | Length 702mm X Width 542mm X Height 574mm |
Refrigerated Reagent Positions | 15 reagent positions, 2 ~ 8℃ | Weight | 65kg |