Haemodialysis Machine WL W-T2008-B

  • W-T2008-B Hemodialysis Machine integrates circuit control, monitoring, blood circulation, and hydraulic systems, for HD and other blood purification in adult chronic renal failure patients (medical use only).

  • Supports hemodialysis, isolated ultrafiltration, etc. Features intelligent dual operation, LCD touch + button interface, backup pump, heparin pump, and 30-minute optional emergency power.

  • Precise A/B ceramic proportion pump; dialysate/temperature/UF rate control; multiple disinfection modes. Monitors blood leak, bubbles, and conductivity with alarms.

Description

Technical Specifications

Configurations:The product is composed of circuit control system, monitoring system, blood extracorporeal circulation control system and hydraulic system, in which W-T6008S includes filter connector, replacement fluid connector, BPM and Bi-cart.

Intended use: 

W-T2008-B Hemodialysis Machine is used for HD dialysis treatment for adult patients with chronic renal failure in Medical Departments

See the instructions for details

Application Purpose of this Device

  • W-T2008-B haemodialysis machine is applicable for chronic renal failure and other blood Purification treatment.
  • This device should be used in medical units.
  • This device is specially designed, produced and sold for renal failure patients to receive haemodialysis, which is not allowed to be used for other purposes.

Forms of Therapy:

Haemodialysis, Isolated Ultrafiltration, Sequential Ultrafiltration, Hemoperfusion, etc.

Features:

  • Intelligent Double Operation System
  • LCD touch screen with button interface
  • Emergency power 30Mins (Optional)
  • Blood Pump
  • Spare Pump (for standby and also can be used for hemoperfustion).
  • Heparin Pump
  • Hydraulic compartment(Balance chamber + UF pump).
  • Operation, Alarm information Memory function.
  • A/B Ceramic proportion pump , High precision, Corrosion-proof, Accuracy.
  • Introduction
  • Size & Weight Size: 380mm×400mm×1380mm (L*W*H)
  • Area: 500*520 mm
  • Weight: 88KG
  • Power Supply AC220V, 50Hz / 60Hz, 10A
  • Input power: 1500W
  • Back-up battery: 30 minutes (optional)
  • Water input pressure: 0.15 MPa ~0.6 MPa
  • 21.75 P.S.I. ~87 P.S.I.
  • Water input temperature: 10℃~30℃
  • Working environment: temperature 10ºC ~30ºC at relative humidity of no more than 70%.

Dialysate

  Dialysate temperature: preset range 34.0℃~39.0℃

  Dialysate flux: 300~800 ml/min

  Dialysate concentration: 12.1 mS/cm ~16.0 ms/cm, ±0.1 ms/cm

  Dialysate mixing ratio: can set variety ratio.

  UF rate Flow range: 0 ml/h ~4000 ml/h

  Resolution ratio: 1ml

  Precision: ±30 ml/h

  Extracorporeal Part

  • Venous pressure: -180 mmHg ~+600 mmHg, ±10 mmHg
  • Arterial pressure : -380 mmHg ~+400 mmHg, ±10 mmHg
  • TMP pressure : -180 mmHg ~+600 mmHg, ±20 mmHg
  • Blood pump flow range: 20 ml/min ~400 ml/min(diameter:Ф6 mm)
  • Spare pump flow range: 30 ml/min ~600 ml/min(diameter:Ф8 mm)
  • Resolution ratio: 1 ml
  • Precision: error range ±10ml or 10% of reading

Heparin Pump

  • Syringe size: 20, 30, 50 ml
  • Flow range: 0 ml/h ~10 ml/h
  • Resolution ratio: 0.1ml
  • Precision: ±5%

Sanitize

  • Hot decalcification
  • Time: about 20 minutes
  • Temperature: 30~60℃, 500ml/min.
  • Chemical disinfection 
  • Time: about 45 minutes
  • Temperature: 30~40℃, 500ml/min.
  • Heat disinfection :
  • Time: about 60 minutes
  • Temperature: >85℃, 300ml/min.
  • Storage Environment Storage temperature should be between 5℃~40℃, at relative humidity of no more than 80%.

Monitoring System

Dialysate temperature: preset range 34.0℃~39.0℃, ±0.5℃

Blood leak detection: Photochromic

Alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate

Bubble detection:ultrasonic

Alarm when a single air bubble volume is more than 200µl at 200ml/min blood flow

Conductivity: acoustic-optic, ±0.5%

Reviews (0)

Reviews

There are no reviews yet.

Be the first to review “Haemodialysis Machine WL W-T2008-B”