Electrolyte Analyzer MTN 6000
-
Auto electrode tracking/calibration (eliminates drift). Direct ISE measurement for K⁺/Na⁺/Cl⁻/Ca⁺⁺/Li/pH in serum, whole blood, CSF, diluted urine.
-
20-position auto-sampler (unlimited loading), 150μL sample. 1min/test, auto sampling/rinsing/printing. Touchscreen + RS232 (connect workstations).
-
Transparent electrode cavity (easy observation). Solenoid valve flushing (anti-clogging), auto protein cleaning. Operates 5-40℃, ≤85%RH, 220V±22V.
Description
Performance characteristics
-
Automatic tracking and calibration of electrode status eliminates the problem of electrode drift. It can directly measure the concentrations of K+, Na+, Cl -, Ca++,Li, and pH in serum, plasma, whole blood, cerebrospinal fluid, and diluted urine.
-
The transparent electrode cavity makes it easier to observe。
-
Sensor based touch screen design makes operation more convenient.
-
Intelligent maintenance free design.
-
Fully automatic sampling, equipped with an automatic sampler, with 20 sample positions for cyclic use and unlimited sample addition; Direct measurement using ion selective electrode method.
Technical parameter
Test project:K+、Na+、Cl-、Ca++、nCa++、TCa++、Li、PH
Measuring method:Ion selective electrode (ISE) direct method.
Sample type:Serum, plasma, whole blood, cerebrospinal fluid, and diluted urine.
Sample size:150μl
Injection method:Fully automatic sampling, equipped with an automatic sampler,with 20 sample positions for cyclic use and unlimited
sample addition.
Washing method:By using an electromagnetic valve for rapid flushing of gas-liquid segments, the electrode pipeline is thoroughly
cleaned and not easily clogged.
Measurement time:Not exceeding 1 minute (depending on the measured equilibrium time, including suction, rinsing, and printing).
Calibration:Automatic tracking, calibration and correction of electrode status, eliminating the problem of electrode drift.
Quality Control:Specially designed quality control testing and statistical functions, easy to use.
Intelligent design:Calibration, injection, measurement, rinsing, display, and printing of reports are automatically completed by the instrument.
Automatic flushing:Automatic scheduled cleaning function for protein solution to remove pipeline protein adsorption.
Display:Large screen sensor touch screen design, fully Chinese display, more convenient operation, and easy conversation between people.
Printer:Built in high-speed low-noise thermal printer.
Interface:Standard RS232 interface, capable of connecting workstations and central computers.
Usage conditions
Power Supply:-220V±22V,50±1Hz
Ambient temperature:5℃-40℃;
Relative humidity:≤85%RH