UV-Vis Spectrophotometer JK 752N

  • Functions & Applications:The 752N UV-Vis spectrophotometer is practical and easy to operate. It has a digital LCD display for readings, automatic blank correction, suitable for teaching and routine analysis.

  • Design & Performance:Manual wavelength setting with light-touch keys for ease of use. Automatic adjustment of 0%T and 100%T. High-performance grating and receiver, low stray light and high accuracy.

  • Parameters & Configuration:Spectral bandwidth 4nm, wavelength 195 – 1020nm. Socket-mounted tungsten lamp for easy replacement. RS232 data interface. Lightweight (weight 8kg).

Description

Technical Specifications

The new 752N UV-Vis spectrophotometer is practical in function and easy to operate. The main unit is equipped with a digital liquid crystal display, enabling intuitive and clear readings. It automatically corrects the blank and enables error-free T/A conversion. Widely used to meet the needs of teaching and routine analysis.

Instrument Functions

  • A: Absorbance testing
  • T: Transmittance testing
  • C: Single standard sample concentration testing
  • F: Concentration testing by the coefficient method

Instrument Features

  • Manual wavelength setting: With light-touch buttons, it is easy to operate and use.
  • Automatic adjustment of 0%T and 100%T: Enables error-free T/A conversion.
  • High-performance grating and receiver: Offers good monochromaticity, high accuracy, and low stray light.
  • Socket-mounted tungsten lamp: Facilitates lamp replacement and debugging.
  • Liquid crystal digital display: Directly displays transmittance, absorbance, and concentration clearly and accurately.

TECHNICAL INDICATORS AND BASIC PARAMETERS

752N

Model

752N

Spectral Bandwidth

4nm

Wavelength Range

195 – 1020nm

Wavelength Accuracy

±2.0nm

Wavelength Repeatability

≤0.5nm

Photometric Range

0 – 200%T, 0.3 – 3A, 0 – 9999C, 0 – 9999F

Photometric Accuracy

±0.002A (0 – 0.5A), ±0.004A (0.5 – 1A), ±0.5%T (0 – 100%T)

Photometric Repeatability

≤0.001A (0 – 0.5A), ≤0.002A (0.5 – 1A), ≤0.2%T (0 – 100%T)

Stray Light

≤0.2%T (220/360nm)

Baseline Drift (Stability)

≤0.002A/h (500nm, 0A)

Noise

±0.2%T (0%T line); ±0.5%T (100%T line); ±0.002A (at 500nm)

Light Source

Long-life tungsten lamp/imported deuterium lamp

Power Supply

AC90 – 250V, 50/60Hz

Dimensions

430×370×200mm

Weight

8kg