Anesthesia Machine PA 01

  • Pneumatically driven electronically controlled system

  • Built-in ventilator, LED displayscreen

  • With 1 patient circuit for adult (Option:patient circuit for children).

  • 4-tube flow meters, 02 & N02 ;0.1 L/Min~ 10 LMin;

Description

Technical Specifications

  • Pneumatically driven electronically controlled system
  • Built-in ventilator, LED displayscreen
  • With 1 patient circuit for adult (Option:patient circuit for children).
  • 4-tube flow meters, O2 & N2O ;0.1 L/Min~ 10 L/Min;
  • One vaporizer,Enflurane/Isoflurane/Sevoflurane; (Option: Halothane)
  • Oxygen Flush: 25~75 L/Min;
  • Respiration Mode: IPPV, SIPPV, SIMV, Manu
  • Tidal Volume: 50~1500 mL;
  • I:E: 1:1.5, 1:2.0,1:2.5, 1:3, 1:4;
  • Respiration Frequency: 4~40 bpm;
  • Ptr: -10~10 hPa;
  • CO2 absorber capacity: 1 kg
  • Alarm parameter: upper /lower limit of airway pressure, tidal volume exceed limitation, no tidal volume output, apnea, AC power malfunction, battery lowpower

Loop part

 
  • Integrated design, integrated breathingcircuit

Evaporator

  • High-precision single evaporator configuration, enflurane or isoflurane or Sevflurane or Haloflurane
  • Concentration adjustment range: 0.5 ~ 5%VOL
  • Evaporator flow, pressure, and automatic temperature compensation function
  • Self-locking feature

Configuration

  • Vaporizer: one vaporizer,Enflurane/Isoflurane/Sevoflurane ( Option: Halothane )
  • Anesthesia gas concentration range:0.5~5%
  • Soda lime absorption system: resistant to high temperature of 176 degrees Celsius.
  • ISO certificates approved
  •  